posted on 2023-05-23, 11:58 authored by Gorjiara, T, Hill, R, Kim, J-H, Kuncic, Z, Adamovics, J, Baldock, CPRESAGE is a dosimeter that is suitable for 3D dosimetry. To be used as an ideal dosimeter, however, it should present radiologically water equivalent properties. In this work, we have investigated the radiological properties of three different PRESAGE® formulations. The radiological water equivalence was assessed by comparing the photon cross sections and radiation dosimetry properties of the three different PRESAGE® formulations with the corresponding values for water. Relative depth doses were calculated using Monte Carlo methods for 75, 125, 180 and 280 kVp and 6 MV x-ray beams. Based on the results of this study, the PRESAGE® formulations with lower halogen content are more dosimetrically water equivalent.
History
Publication title
6th International Conference on 3D Radiation Dosimetry (IC3DDose 2010), South Carolina, USAVolume
250Pagination
244-248ISSN
1742-6588Department/School
Research ServicesPublisher
IOP Publishing LtdEvent title
6th International Conference on 3D Radiation Dosimetry, IC3DDose 2010Event Venue
Journal of Physics: Conference Series, 250 (1)Date of Event (Start Date)
2010-08-22Date of Event (End Date)
2010-08-26Rights statement
Copyright 2010 IOP Publishing Ltd.Repository Status
- Restricted