posted on 2023-05-18, 13:25 authored by Clark, TD, Eliason, EJ, Sandblom, E, Hinch, SG, Farrell, AP
The HemoCue haemoglobin analyser consistently overestimated haemoglobin concentration ([Hb]) in the blood of all fish species (sockeye salmon Oncorhynchus nerka, Chinook salmon Oncorhynchus tshawytscha, Pacific bluefin tuna Thunnus orientalis and chub mackerel Scomber japonicus) by 22–50% (9·9–36·0 g l−1) over a [Hb] range of 20–160 g l−1. The systematic nature of the overestimation, however, allowed the formulation of an accurate calibration equation that can be used to correct values of [Hb] measured by the HemoCue in field studies.