Fast pulsed amperometric waveform for miniaturised flow-through electrochemical detection: Application in monitoring graphene oxide reduction

journal contribution
posted on 2023-05-20, 07:49 authored by Islam, MA, Koreshkova, AN, Vipul GuptaVipul Gupta, Trevor LewisTrevor Lewis, Miroslav MackaMiroslav Macka, Brett PaullBrett Paull, Mahbub, P
A fast pulsed waveform (2-steps at +400 mV and −400 mV, 10 ms each, 50 Hz cycle repetition) was developed incorporated within a miniaturised EC detector, and used to demonstrate the pulsed amperometric detection (PAD) of hydrazine in a flow-based analytical system. In contrast to standard amperometric detection (AD), a subsequent surface oxide reduction step was applied within the PAD waveform. The PAD applied within a simple flow-injection analysis (FIA) method provided a limit of detection (LOD) just as low as 0.8 nM of hydrazine (RSD 5%, n = 9, linearity r2 = 0.99 for 1 nM-100 μM), which is one order of magnitude lower than the lowest LOD reported todate. Significantly, the LOD was obtained using 5–20 times lower electrode surface area, flow rate, and sample volume, than in previous methods, with 15 times faster cycle repetition frequency than alternative PAD based studies reported in literature. Practical application of the PAD waveform was demonstrated by monitoring the temporal consumption of reducing agents during graphene oxide reduction experiments, separated using chromatographic methods (hydrazine in IC and ascorbic acid in HPLC).

History

Publication title

Electrochimica Acta

Volume

328

Article number

135087

Number

135087

Pagination

1-8

ISSN

0013-4686

Department/School

School of Natural Sciences

Publisher

Elsevier

Place of publication

Oxford, England

Rights statement

© 2019 Elsevier Ltd. All rights reserved.

Repository Status

  • Restricted

Socio-economic Objectives

Industrial chemicals and related products not elsewhere classified; Scientific instruments