University Of Tasmania
Browse
245 Breadmore - Anal Chem 73 (2001) 0820 - 0828.pdf (169.94 kB)

On-Capillary Ion-Exchange Preconcentation of Inorganic Anions in Open-Tubular Capillary Electrochromatography with Elution Using Transient-Isotachophoretic Gradients. 2. Characterization of the Isotachophoretic Gradient

Download (169.94 kB)
journal contribution
posted on 2023-05-16, 12:57 authored by Michael BreadmoreMichael Breadmore, Miroslav MackaMiroslav Macka, Avdalovic, N, Paul HaddadPaul Haddad
Diffuse transient-isotachophoretic boundaries can be used as an elution gradient of increasing eluotropic strength to elute inorganic anions that have been preconcentrated on an open-tubular ion-exchange stationary phase prior to electrophoretic separation. The generation and characteristics of these gradients for elution after preconcentration have been investigated. The gradients are generated by placing a low-mobility, weak ion-exchange competing anion in the capillary (weak electrolyte, WE), and a high-mobility, strong ion-exchange competing anion in the electrolyte vials (strong electrolyte, SE). Application of voltage establishes a diffuse boundary with the composition changing from the weak anion to the strong anion. Comparison of elution gradients generated with different electrolyte systems was accomplished by comparing the eluotropic strength (a function of eluent concentration, ion-exchange selectivity coefficient, and charge) and the shape of the profile as it changes from WE to SE. The ion-exchange selectivity coefficient of the SE competing anion is important in establishing a sharp change in elution strength. A large difference in mobility between the WE and SE competing anions gives an SE with a higher final eluotropic strength, but the slope of the gradient is shallower. This results in a reduction in the efficiency of analyte focusing. To ensure maximum focusing efficiency, the WE and SE electrolytes should be selected such that the SE has the highest possible eluotropic strength for a given concentration of WE. The SE competing anion should also have a sufficiently low electrophoretic mobility to ensure focusing for the maximum number of analytes, and the mobility difference between the WE and SE competing anions should be as small as possible.

History

Publication title

Analytical Chemistry

Volume

73

Issue

4

Pagination

820-828

ISSN

0003-2700

Department/School

School of Natural Sciences

Publisher

American Chemical Society

Place of publication

Washington DC

Rights statement

Copyright © 2001 American Chemical Society- Reproduced by permission of The Royal Society of Chemistry (RSC)

Repository Status

  • Open

Socio-economic Objectives

Expanding knowledge in the chemical sciences

Usage metrics

    University Of Tasmania

    Exports