Selectivity tuning via temperature pulsing using low thermal mass liquid chromatography and monolithic columns
journal contribution
posted on 2023-05-17, 22:01authored byPursch, M, Eckerle, P, Gu, B, Luong, J, Cortes, HJ
Low thermal mass LC was applied to the capillary LC separation of a complex insecticide mixture by increasing temperature and decreasing gradients, as well as fast selected temperature pulses to increase resolution of overlapped components. The technology was applied using a new generation of capillary monolithic stationary phases. Considerable peak shifts and selectivity changes were observed for given temperature conditions. The concept of temperature pulsing during an elution profile shows promise for increasing resolution in difficult separations and can provide a relatively simple means to solve coelution problems.