PURPOSE: The aim was to investigate the stability of cefazolin in elastomeric infusion devices.
METHODS: Elastomeric devices (Infusor LV) that contain cefazolin (3 g/240 mL and 6 g/240 mL) were prepared and stored at 4°C for 72 hours and then at 35°C for 12 hours, followed by 25°C for 12 hours. An aliquot was withdrawn at predefined time points and analyzed for the concentration of cefazolin. Samples were also assessed for changes in pH, solution color, and particle content.
FINDINGS: Cefazolin retained acceptable chemical and physical stability over the studied storage period and conditions.
IMPLICATIONS: These findings will allow the administration of cefazolin by the Infusor LV elastomeric device in the outpatient and remote settings.