The effect of physiologic and mechanical aging on the performance of peak flowmeters
journal contribution
posted on 2023-05-16, 16:59authored byJohns, DP, Side, E, Kendrick, AH, Williams, TJ, Eugene WaltersEugene Walters
Purpose: To investigate the effects of physiologic and mechanical aging on peak flowmeters. Materials and methods: Eight each of MiniWright (MW; Clement Clark; Harlow, UK), Personal-Best (PB; HealthScan Products; Cedar Grove, NJ), Vitalograph (V; Vitalograph Ltd; Buckingham, UK), and Breath- Taker (BT; Medical Development Australia; Melbourne, Australia) peak flowmeters were assessed for accuracy and repeatability before and after aging using a computer-driven syringe to deliver peak flows from 100 to 700 L/min. Four of each type of flowmeter were physiologically aged by normal subjects performing up to six peak flows daily for 1 year. The remaining four of each flowmeter were mechanically aged using an accelerated aging device to deliver 2,000 exponential waveforms with a peak flow of 600 L/min over a period of 3 h. Results: The V and BT flowmeters were linear and accurate over the range 100 to 700 L/min, while the PB overread at high flows. The MW was alinear throughout. The SD of the difference between readings before and after aging ranged from 8.6 to 40.6 L/min (mean, 9.2). Comparing the slopes of the relationship of actual against reference peak expiratory flow (PEF) showed that 16 flowmeters-5 BTs, 6 MWs, 4 PBs, and 1 V had no significant change in slope after aging. Mechanical aging caused a consistent underreading in PEF at high flow rates. Physiologic aging showed a more variable pattern both within and between flowmeter types. The MW was the most affected by physiologic aging, producing overestimates of PEF by as much as 100 L/min at 500 L/min. Conclusions: We conclude that the effects of physiologic and mechanical aging are different, and that while mechanical aging may provide a guide to the effects of aging, studies using physiologic aging would be more appropriate.